testo 162-T2: Set Up Two-Point NTC Temperature Monitoring

Quick answer: The testo 162 T2 records temperature from two compatible external NTC probes, with a local display and supported Wi-Fi/cloud monitoring. Connect and label the probes, place them at the two required locations, configure two independent channels and verify the reports and alarms. It has no relative-humidity sensor.

How the T2 differs from the T1 and H series

Testo identifies the T2 as product 0572 1622, with two connections for external NTC temperature probes. The T1 has one integrated temperature sensor instead; the H1 and H2 address temperature plus RH. The T2’s external probes allow the sensor points to be inside a cabinet or separated physically from the display and Wi-Fi body.

Testo lists NTC accessories such as penetration probes with flat ribbon cables, long-cable probes and surface probes. Choose a sensor based on the medium and mounting method; a surface probe on a pipe is not equivalent to a liquid-core reading. For general guidance on choosing compatible sensors, see Temperature Probes Explained.

Place and identify two probes

  1. Define Point A and Point B, including what each temperature should represent.
  2. Select compatible NTC probes, checking measuring range, protective construction, immersion depth and cable route.
  3. Connect both probes to the correct T2 inputs and label channel names to match the physical positions.
  4. Mount the logger securely and enroll it in the supported Wi-Fi monitoring service using the current commissioning workflow.
  5. Set logging intervals and separate alarm rules where the process requires different limits.
  6. Compare recorded times, probe identity and plausible readings, then carry out a controlled alarm/reporting test.

Worked example: refrigerator air and product simulator

Hypothetical: Channel A in refrigerator air reads 3.2 °C, while Channel B in a validated product simulator reads 5.0 °C. The temperature difference is 1.8 °C. The simulator may respond more slowly to door openings; neither reading should be relabeled as the other measurement point. Product release limits require the applicable approved protocol and sensor arrangement.

Measurement positions and cable issues

Reading patternPotential explanationNext check
Both channels shift simultaneouslyReal chamber change or common environmental effectConfirm room events and logger timestamps
Only Channel B jumpsProbe contact, route or local thermal eventCheck B mounting and connector
One channel missingProbe absent, damaged or disconnectedVerify input, sensor type and cable
Large stable differenceDifferent media/locations or sensor offsetsCo-locate and compare against documented references

Data monitoring service and alarms

Current Testo US product descriptions specify the online monitoring ecosystem, wireless LAN, app setup and required Data Monitoring license. Check service compatibility with older Saveris 2 accounts before commissioning. A live reading on the local display does not mean that a remote alert has been delivered; prove both separately. For a comparable logger with a built-in sensor, see the testo 162-T1 setup guide.

Limits: no humidity and no automatic compliance

The two NTC inputs measure temperature, not humidity, and do not turn an air measurement into a direct product-core reading. Preserve sensor serial details, positioning records, alarm tests and the applicable calibration program. Select a testo 162 H model if temperature/RH monitoring is the real objective. If the monitoring requirement includes humidity with an external probe, see the testo 162-H2 configuration guide.

Official documentation and references